Douglas O. S. Wood
Please contact Theresa McBride (tmcbride@nrao.edu) for a paper copy of this memo, which includes figures of the front end electronics and some examples of preliminary results obtained with the 7 mm system early in 1994.
To date, the VLA has been outfitted with nine Q band systems. Their design is similar to the Q band systems of the VLBA except for some modifications to the polarizers, mixers and isolators. Obvious changes were necessary to accommodate the VLA local oscillators and the different secondary reflectors used on the VLA. The LO scheme makes use of some components of the existing X band system (F12 frequency converter). The LO is provided by tripling the F3 module.
The feeds are mounted 66 inches above the vertex house at a nearly horizontal position. This location has the advantage that if a significant droop in the antenna structure degrades high frequency performance at low elevations, the droop could be corrected (in part) by a compensating rotation of the subreflector. Such a compensation is not in use today but may be used in the future.
The antennae selected for the upgrade were: 3,4,8,12,14,16,22,25, 27 and 13 chiefly because these antennae had higher than average K-band efficiency and better than average pointing. A tenth system should be installed by June 1994. It will be used as a functional spare and will be removed from service if necessary.
The construction was completed on time and on budget according to the following
schedule:
For an interferometer of N elements, the rms in an image is given by
Further tests are underway to investigate high frequency (49 GHz) performance which will be lower. At 49 GHz we can expect approximately twice the system temperature, twice the attenuation of the source due to the atmosphere. The combination of these effects with an expected loss of ~30% in aperture efficiency predicts a signal-to-noise at 49 GHz which will be ~20% that of the nominal operating frequency of 43.3 GHz. Tests presently underway will explore the performance of the high frequency end of the band further.
Because good observing conditions are more rare at mm than at cm wavelengths, observers are encouraged to pick sources that are up at night and to schedule their subarray (non-Q band) observations at a lower frequency (X band or below). Phase calibration at 7 mm should be much more frequent than at other VLA bands, perhaps as often as every 5 minutes. If weather conditions prohibit successful Q band observations, one may want to use the entire array at X band or some lower frequency instead.
If a target source contains a maser source (e.g. SiO), you should exploit the technique of "Phase Referencing" in which phase errors are determined by a maser observation in one IF and applied to the data taken in the other IF. This can be done either in line or continuum mode. In continuum mode one uses a very narrow bandwidth centered on the maser in one IF and a 50 MHz bandwidth for the source continuum observation in the other IF. Some of the first VLA observations at Q band have used this technique but as of this date these data have not been reduced. Phase referencing with a maser source in the field should improve images and more frequent phase calibration (every 5-10 min) would help. Under some conditions 360 degree phase variations over 3 to 4 minutes are seen, and Q band observations are probably not possible.
Reference pointing scans are made in interferometer pointing mode (IR) and take a minimum of 3 minutes (10 sec integrations). They are usually made with the full array at X band where signal-to-noise is highest. It is important to select reference pointing calibrators to be as close as possible to your source both spatially and temporally. Reference pointing scans should be made approximately every hour and should not be applied to sources more than 30 degrees away in AZ or EL from the pointing calibrator. Usually the phase calibrator is a good choice for pointing calibrator. If possible, select a reference pointing calibrator with a smaller RA than your program source. This way the program source will drift through the position of the reference pointing scan during the program source observation. Avoid observing sources within 10 degrees of the zenith where changes in AZ are too rapid to calibrate. The results of each reference pointing scan are printed on hard copy at the site. If you are not present for your observations, ask the operator to send the output to you. It can be a helpful diagnostic of your run.
3C286 @ 43.3 GHz = 1.86 Jy
3C48 @ 43.3 GHz = 0.53 Jy
Other values can be obtained from the polynomial fits given by Ott et al. Longer integrations are required (10 to 15 min) especially on 3C48, to get a good flux calibration. We may use some planets for flux calibration but the procedure for this is still being developed.
Be aware that there are several restrictions as to what you can do in the main
and subarrays. Ken Sowinski has written a memo which describes the details.
The basic rules to follow for your main and sub array files are:
(1) the
main and subarrays must be in the same mode (line or continuum)
(2) they
must use the same integration time
(3) they cannot perform reference
pointing scans at the same time
(4) they must use the same band width in
each IF (AC and BD can have different bandwidth but the main and subarrays
must have the same bandwidth configuration.)
Your main and subarray files must be prepared independently in OBSERVE and it is up to you to check them for any conflict. Contact Ken (ksowinsk@nrao.edu) if you have questions that are not answered in his memo.
Such an expansion, however, will require more than simply outfitting the remaining antennae with Q band receivers. The first 9 systems were chosen because they were our best performers in terms of pointing and aperture efficiency. In order for the remaining antennae to be most useful in a full 28 element 7 mm system, their pointing and surface accuracies must be improved. It is likely that surface accuracy can be improved with only a man-power effort (no equipment or capital is needed). An example of this is antenna 4 (see above). Improving the pointing is another matter and may require replacement or overhaul of major parts such as the AZ and/or EL bearings. Reference pointing may reduce these problems to a manageable level with out significant hardware cost. The discussion of this is beyond the scope of this memo, but it is an upgrade that would improve the performance not only at 7 mm but for all VLA bands.
Second, the available number of channels in the correlator, especially for broad band observations needs to be increased. With only 8 or 16 channels at 50 MHz, line profiles are not well determined. A correlator with a greater number of channels would be a significant improvement.
Third, obtaining a broad band continuum channel (1 GHz, say) while in spectral line mode might make self-calibration on weaker line sources possible. Even without the advantage of self-cal, having a high sensitivity continuum image would be and important advantage.